Xuanlin Machinery is a company of producing CNC Routers for more 15 years and we have customers worldwidely..For all parts of machine ,we can offer for customers ,such as spindles ,water cooling and air cooling,drivers and motors -- servo motor and driver ,stepper motor and driver, all models that CNC Router Machine used. Variable Frequency drivers, for from 2.2 kw to 11 kw, square guide blocks , cnc bits-- wood bit ,V shaped, single flute,double flute, oil system -- manual oil system and auto oil system, dust collectors-- single bag dust collector, double bags dust collector, from 2.2 kw to 5.5 kw.We have all what you want. Woodworking Machinery Spare Parts,Delta Table Saw Parts,Delta Band Saw Parts,Delta Saw Parts JINAN XUANLIN MACHINERY CO., LTD. , https://www.cncfiberlaser.com
1. Turn on the screen to display half screen or crash, how to solve it?
Answer: Restart the calibrator, or take out the calibrator's battery, leave it for about 30 seconds, then put the battery into the calibrator and reboot.
2. After booting up, it shows the non-stop reset, how to solve
Answer: Please replace the battery.
3. After the power button is pressed, it can be started normally. After the power button is released, the display screen is black. How to solve it?
Answer: You need to return to the factory for repair.
4. Can't communicate with PC (or pressure module; temperature module), how to solve it?
Answer: Enter the maintenance mode of the calibrator and set the communication mode.
5. When using the current function of the multimeter to detect the current output function of the calibrator, the multimeter has no reading. How to solve it?
Answer: a. Check if the line is connected correctly; if the function is set correctly;
b. Check insurance;
c. Check if the 24V power supply is normal. If it is not normal, you need to return to the factory for repair.
6. When the calibrator is set in the analog transmitter mode, there is no current output. How to solve it?
Answer: a. Check if the line is connected correctly; if the function is set correctly;
b. Check insurance.
7. When the calibrator is set to measure the current function, the current cannot be measured. How to solve it?
Answer: a. Check if the line is connected correctly; if the function is set correctly;
b. Check insurance.
8. When connecting the two-wire transmitter with the calibrator, there is no current output in the loop. How to solve it?
Answer: a. Check if the line is connected correctly; if the function is set correctly;
b. Check insurance;
c. Whether the 24V loop power supply is turned on;
d. Check whether the 24V loop power supply meets the power requirements required by the object under test. If not, use a regulated source or other device to power the transmitter.
9. There is no output of the resistor (voltage). How to solve it?
Answer: a. Check if the line is connected correctly; if the function is set correctly;
b. Check insurance.
10. The resistance output is incorrect. How to solve it?
Answer: a. Check if the line is connected correctly; if the function is set correctly;
b. Check if the excitation current is within the range of the excitation current of the selected resistance range.